Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter "Overview of validated
procedures".
During sterilization, do not exceed 138 °C.
Do not sterilize in hot air.
Do not sterilize with gamma radiation.
Do no sterilize with STERRAD ®!
7.2
Reprocessing limitation
Frequent reprocessing can damage this product. Observe the
safety instructions for checking the product. With visible damage
or functional impairment, the product may no longer be used.
7.3
Dismantling
The product must be prepared while it is dismantled.
1. Pull the cable from the handle (no picture).
3
2. Hold the handle (4) so the movable grippers are not blocked.
Press the button and pull the shaft tube (3) out of the handle.
WARNING! In order to prevent damage to the instrument, do not
grip the instrument by the jaws.
4