Check; Packaging - Erbe LAP BiSect Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 56
OBJ_DOKU-184279-008.fm Seite 34 Montag, 5. November 2018 8:20 08
4. Start a tested program with the following properties:
– Sufficient cleaning effect.
– Thermal disinfection: 5 to 10 minutes at 90 to 95 °C,
A
≥ 3000.
0
– Final rinse with distilled or deionized water.
– Sufficient product drying (no visible fluid residues).
5. If contaminants are still visible on the product after the end
of the routine, repeat precleaning and machine cleaning and
disinfection.
6. Remove the irrigation tubes from the insulation tube and the
shaft.
7.7

Check

1. Check the product for visible damage and wear:
– Damage to the product, e.g. cracks, rough surface, splin-
tering.
– Damage to the product insulation and/or cable/connec-
tor insulation, e.g. cracks and breaks.
– Bending of the product or parts of the product. Do not at-
tempt to straighten bent instruments!
2. Check the mechanical functionality of the product:
– Can the jaws be opened and closed?
– Can the handle be opened and closed?
If damaged, do not use this product!
3. Oil all moveable parts of the handle and the joint area of the
insert using a physiologically acceptable oil (e.g. paraffin oil).
7.8

Packaging

1. Pack the product in disposable sterilization packaging (single
or double packaging) made from paper/foil and/or in a ster-
ilization container.
34

Publicidad

Tabla de contenido
loading

Tabla de contenido