Descargar Imprimir esta página

Henry Schein HS IV Warmer Mode D'eMploi página 4

Calentador de fluidos para infusión

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 6
XII. Package and Transportation
In order to guarantee safe transportation, the unit is packaged in a special carton package of with EPE foam. The package can resists
humidity, rain and fog.
The package is printed with the following logos:
No.
Logo Name
1
Up
2
Handle with care
5
No humidity
7
No rolling
The size and weight of the package are printed on the carton.
Before delivery, the products are categorised and packed according to their functions, weights, volumes and vulnerability, so the package
can meet safety requirements of transportation and storage. There is a detailed packing list in each package.
XII. Installation
After removal from packaging the user can carry out all procedures in the Section VIII and IX,
XIII. Symbol and Mark
No.
Position
1
Earthing Terminal
2
Name Plate
3
Name Plate
4
Name Plate & Documents
Appendix
The Infusion Fluid Warmer is intended for use in the electromagnetic environment specified below. The customer or the user of the product should ensure
that it is used in such an environment.
Emissions test
CISPR11 RF emissions
CISPR11 RF emissions
IEC61000-3-2 Harmonic
emission
IEC61000-3-3Voltage
fluctuations / flicker emissions
Guidance and manufacturer's declaration – electromagnetic immunity
IMMUNITY test
Electrostatic discharge
(ESD)IEC61000-4-2
Electrical fast transient/burst
IEC61000-4-4
Surge
IEC61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC61000-4-11
Power frequency (50/60 Hz)
magnetic field
IEC61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.
Logo Mark
Symbol and Mark
Guidance and manufacturer's declaration—electromagnetic emissions
Compliance
Electromagnetic environment -- guidance
The Infusion Fluid Warmer uses RF energy only for its internal function.
Group 1
Therefore, its RF emissions are very low and unlikely to cause any interference in nearby electronic equipment.
Class B
The Infusion Fluid Warmer is suitable for use in all establishments, including domestic establishments that
Not applicable
are directly connected to a mains power supply.
Not applicable
IEC60601
Test level
Compliance level
±6kV contact
±6kV contact
±8kV air
±8kV air
±2kV for power
±2kV for power
supply lines
supply lines
±1 kV line(s) to line(s)
±1 kV line(s) to line(s)
±2 kV line(s) to earth
±2 kV line(s) to earth
<5 % U T (>95 % dip in U T ) <5 % U T (>95 % dip in U T ) Mains power quality should be that of a typical commercial or hospital
for 0,5 cycle
for 0,5 cycle
40 % U T (60 % dip in U T ) 40 % U T (60 % dip in U T ) during power mains interruptions, it is recommended that the warmer
for 0,5 cycle
for 5 cycles
70 % U T (30 % dip in UT) 70 % U T (30 % dip in U T )
for 25 cycles
for 25 cycles
<5 % U T (>95 % dip in U T ) <5 % U T (>95 % dip in U T )
for 5 s
for 5 s
3A/m
3A/m
Description
Earth Protection
Class I Type B
Disposal of waste products must follow
local laws and regulations
Notice: please see the manual
Electromagnetic environment--guidance
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material , the relative humidity
should be at least 30 %.
Mains power quality should be that of a typical commercial
or hospital environment.
environment. If the user of the warmer requires continued operation
be powered from an uninterruptible power supply.
Power frequency magnetic fields should be at levels characteristic of a
typical location in a typical commercial or hospital environment.
4

Publicidad

loading

Este manual también es adecuado para:

9882043