Step-by-Step Instructions
6.5

Searching for the Apex

6.5.1

Coronal and Medial Zone

Slowly introduce the file into the canal and advance the file with slow
clockwise turns. The cursor on the tooth icon indicates the progression
of the file inside the canal.
Further progression of the file in the canal is shown both by the cursor
on the tooth icon and by numerical value on the graphical scale.
ProPex
progressive rate beeps. If the bar graph makes sudden large
movements in the upper part of the canal, continue slightly towards the
apex so the signal returns to normal.
The scale indication on the ProPex
distinct length or distance in mm or other linear units. It simply
indicates the file progression towards the apex.
B EN PIIL DFU WEB / Rev.07 / 02-2018 (Old ZF 190292.EN)
®
II emits audible information of file progression via a series of
®
II screen does not represent a
15/48