Warranty; Disposal Of The Product - Dentsply Sirona ProPex II Manual Del Usuario

Ocultar thumbs Ver también para ProPex II:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 147

Warranty

Problem
Display reaction is over-
sensitive: apex/over is
8.
activated before it is
reached.
11
WARRANTY
ProPex
accessories (cables, battery etc.) are warranted for 6 months from the
date of purchase.
The warranty is valid for normal usage conditions. Any modification or
accidental damage will render the warranty void.
12

DISPOSAL OF THE PRODUCT

PLEASE DO NOT THROW AWAY !
This product and all its components must absolutely be recycled
through your supplier.
B EN PIIL DFU WEB / Rev.07 / 02-2018 (Old ZF 190292.EN)
Possible cause
Short circuit due to excess liquid
(irrigation solution, saliva, blood)
in the pulp chamber.
A direct contact of the file with
the gingiva or gingival
proliferations, e.g. a fractured
metal crown.
A direct contact of the file with
metal restorations (crown,
parapulpal post, amalgam
filling).
®
II is warranted for 24 months from the date of purchase. The
Solution
Dry the access cavity with a
cotton pellet / air-blower. In case
of excess bleeding wait until it
has stopped.
For isolation:
- adequate preparation filling;
- use a rubber dam.
Isolate the file by placing 2-3
silicone stoppers on it or insert
the file in a small polyvinyl tube
before use.
31/48

Publicidad

Tabla de contenido
loading

Tabla de contenido