Cleaning, Disinfection and sterilization
N°
Operation
7.
Packaging
8.
Sterilization
9.
Storage
26/48
Operating mode
Pack the devices in "Sterilization
pouches".
Steam sterilization at: 134°C
(237°F) during 3 min.
Keep devices in sterilization
packaging in a dry and clean
environment.
B EN PIIL DFU WEB / Rev.07 / 02-2018 (Old ZF 190292.EN)
Warning
Check the validity period of the
pouch given by the manufacturer
to determine the shelf life.
Use packaging which are
resistant up to a temperature of
141°C (286°F) and in
accordance with EN ISO 11607.
The accessories (lip clip, hook
and fork) must be sterilized
according to the packaging
labelling.
Use fractionated vacuum or
gravity (less preferred)
autoclaves (according to
EN 13060, EN 285).
Use validated sterilization
procedure according to
ISO 17665-1.
Respect maintenance procedure
of the autoclave device given by
the manufacturer.
Use only the listed sterilization
procedures.
Sterility cannot be guaranteed if
packaging is open, damaged or
wet (check the packaging before
using the instruments).