Guidance and Manufacturer's declaration - Electromagnetic Immunity
The Product is intended for use in the electromagnetic environment specified below.
The customer or the user of the Product should assure that it is used in such an
environment.
Immunity
tests
Electrostatic
± 6 kV contact
discharge
(ESD) IEC
± 8 kV air
61000-4-2
Power
frequency
(50/60 Hz)
3 A/m
magnetic
field IEC
61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
B EN PIIL DFU WEB / Rev.07 / 02-2018 (Old ZF 190292.EN)
IEC 60601
Compliance level
Test level
± 6 kV contact
± 8 kV air
3 A/m
Electromagnetic
environment - guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Power frequency
magnetic fields should be
at levels characteristic of
a typical location in a
typical commercial or
hospital environment.