Maintenance
4.1 Calibration
The calibration process is essential in producing precise 3D models.
You should perform the calibration process periodically.
Calibration is required when :
The quality of the 3D model is not reliable or accurate as compared to previous
results.
Environmental conditions such as temperature has changed.
Calibration period has expired.
You can set the calibration period as detailed in Menu > Setting > Calibration
Period(Days)
The calibration panel is delicate component. Do not touch the panel directly.
When the calibration process does not perform properly, check the panel. If
the calibration panel is contaminated, please contact your service provider.
We recommend performing the calibration process periodically.
You can set the calibration period via Menu > Setting > Calibration
Period(Days).
The default calibration period is set to 14 days.