Safety Guide
If the i500 unit is dropped on the floor or if the unit is impacted, it must be
calibrated before use. If the instrument is unable to connect to the software, consult
the manufacturer or authorized resellers.
When not in use, the i500 should be kept mounted on a desk stand or wall mount
stand.
Do not install the desk stand on an inclined surface.
Carefully arrange all cables so that you or your patient do not trip or get caught in
the cable. Any pulling tension on the cables may cause damage to the i500 body.

5.4 Explosion Hazards

WARNING
he i500 system is not designed to be used near flammable liquids or gases, or in
environments with high oxygen concentrations.
There is a risk of explosion if you use the i500 system near flammable anesthetics.

5.5 Electrical Safety

WARNING
The i500 system is a Class 1 device.
To prevent electric shock, the i500 system must only be connected to a power
source with a protective earth connection. If you are unable to insert the i500-
supplied plug into the main outlet, contact a qualified electrician to replace the
plug or outlet. Do not try to circumvent these safety guidelines.
The i500 system only uses RF energy internally. The amount of RF radiation is low
and does not interfere with surrounding electromagnetic radiation.
There is a risk of electric shock if you attempt to access the inside of the i500 system.
Only qualified service personnel should access the system.
Do not connect the i500 system to a regular power strip or extension cord as
these connections are not as safe as grounded outlets. Failure to adhere to these
safety guidelines may result in the following hazards:
The total short circuit current of all connected equipment may exceed the limit
»
specified in EN / IEC 60601-1.
The impedance of the ground connection may exceed the limit specified in EN /
»
IEC 60601-1.
Do not place liquids such as beverages near the i500 system and avoid spilling liquid
on the system.
Condensation due to changes in temperature or humidity can cause moisture
buildup inside the i500 unit, which may damage the system. Before connecting the
i500 system to a power supply, be sure to keep the i500 at room temperature for at
least 2 hours to prevent condensation. If condensation is visible on the product
surface, the i500 should be left at room temperature for more than 8 hours.
You should only disconnect the i500 system from the power supply via its power
cord.
The radiation characteristics of the i500 system makes it suitable for use in industry
and hospitals. (CISPR 11 class A). If the i500 system is used in a residential
environment (CISPR 11 class B), it may not provide adequate protection from radio
frequency communications.
Before disconnecting the power cord, make sure to turn off the power on the device
using the power switch on the main unit.
Only use the power adaptor supplied together with the i500. Using other power
adaptors may result in damage to the system.
CAUTION
Avoid pulling on the communication cables, power cables, etc. used in the i500 system.
Intraoral Scanner i500
17