RATED power frequency magnetic fields (50/60Hz) IEC 61000-4-8
RATED power frequency magnetic fields (50/60Hz) IEC 61000-4-8 Power frequency
magnetic fields should be at levels characteristic of a location in a typical
commercial or hospital environment.
IEC 60601-1-2 Test Level
30 A/m
Conducted RF IEC 61000-4-6
Portable and mobile RF communications equipment, including cables, should
be used no closer to any part of the EUT than the recommended separation
distance as calculated using the equation below, according to the frequency of
the transmitter.
Recommended separation distance
d = [
3,5
]√P
V
1
d = [
3,5
]√P
80 MHz to 800 MHz
E
1
d = [
7
]√P
800 MHz to 2.5 GHz
E
1
where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic
site survey, should be less than the compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with the following
symbol:
Compliance Level
30 A/m
Radiated RF IEC 61000-4-3
Portable and mobile RF communications equipment, including cables, should
be used no closer to any part of the EUT than the recommended separation
distance as calculated using the equation below, according to the frequency of
the transmitter.
IEC 60601-1-2 Test Level
3 V 0,15 MHz – 80 MHz 6 V in ISM
bands between 0,15 MHz and
80 MHz 80 % AM at 1 kHz
Recommended separation distance
d = [
3,5
]√P
V
1
3,5
d = [
]√P
80 MHz to 800 MHz
E
1
d = [
7
]√P
800 MHz to 2.5 GHz
E
1
Compliance Level
3 V 0,15 MHz – 80 MHz 6 V in ISM
bands between 0,15 MHz and 80 MHz 80 %
AM at 1 kHz
Intraoral Scanner i500
21