Maintenance
4.4 Updates to Image Acquisition Software
The Image acquisition software automatically checks for updates when the
software is running.
When a new version of the software is available, the system will automatically
download the new version.
5

Safety Guide

Please adhere to all safety precautions as detailed in this user guide to prevent
human injury and equipment damage. This document uses the words WARNING
and CAUTION when highlighting precautionary messages.
The i500 system should only be operated by dental professionals and technicians
who are trained to use the system. Using the i500 system for any purpose other
than its intended usage as outlined in section "2.1 Intended Use" may result in
injury or damage to equipment. Please handle the i500 system according to the
guidelines in the safety guide.
WARNING
Please read the safety guide carefully. Failure to follow the safety precautions as
highlighted in the guide may result in injury or damage to equipment.
5.1 System Basics
CAUTION
The USB 3.0 cable connector to the Power Hub is the same as a regular USB cable
connector. However, the device may not operate normally unless the regular 3.0
USB cable is used with the i500.
The connector provided by the Power Hub is designed specifically for the i500 and
should not be used with any other device.
If the product has been stored in a cold environment, give it time to adjust to the
temperature of the environment before use. If used immediately, condensation may
occur which may damage the electronic parts inside the unit.
The calibration process should be carried out if the temperature of the environment
has changed.
The calibration panel is delicate. Do not touch the panel directly.
When the calibration process does not perform properly, check the panel. If the
calibration panel is contaminated, please contact your service provider.
Before using the system, check that there are no issues such as physical damage,
loose parts, and wear. If there is any visible damage, do not use the product and
contact the manufacturer or your local representative.
Check the i500 body and its accessories for any sharp edges.
Always keep an eye on the product and your patient while using the product to
for abnormalities.
If the equipment fails to operate normally, such as having issues with accuracy, stop
using the product and contact the manufacturer or authorized resellers.