Electro – Magnetic Compatibility Information
6.2 Elector-Magnetic Immunity
This EUT is intended for use in the electromagnetic environment as specified
below. The customer or the user of the i500 system should ensure that it is used
in such an environment.
Electrostatic discharge (ESD) IEC 61000-4-2
Floors should be of wood, concrete or ceramic tiles. If floors are covered with a
synthetic material, the relative humidity should be at least 30%.
IEC 60601-1-2 Test Level
± 8 kV contact ± 2 kV, ± 4 kV, ± 8 kV,
± 15 kV air
Electrical fast transient/burst IEC 61000-4-4
Mains power quality should be that of a typical commercial or hospital
environment.
IEC 60601-1-2 Test Level
± 2 kV 100 kHz repetition frequency
Surge Line-to-line IEC 61000-4-5
Mains power quality should be that of a typical commercial or hospital
environment.
IEC 60601-1-2 Test Level
± 0.5 kV , ± 1 kV
20
User Manual
Compliance Level
± 8 kV contact ± 15 kV air
Compliance Level
± 2 kV 100 kHz repetition frequency
Compliance Level
± 1 kV
Surge Line-to-ground IEC 61000-4-5
Mains power quality should be that of a typical commercial or hospital
environment.
IEC 60601-1-2 Test Level
± 0.5 kV, ± 1 kV, ± 2 kV
Voltage dips IEC 61000-4-11
Mains power quality should be that of a typical commercial or hospital
environment. If the user of the EUT image intensifier requires continued operation
during power mains interruptions, it is recommended that the EUT image
intensifier be powered from an uninterruptible power supply or a battery.
IEC 60601-1-2 Test Level
0 % UT; 0.5 cycle At 0°, 45°, 90°, 135°,
180°, 225°, 270° and 315°
0 % UT; 1 cycle and 70 % UT; 25/30
cycles Single phase: at 0°
Voltage interruptions IEC 61000-4-11
Mains power quality should be that of a typical commercial or hospital
environment. If the user of the EUT image intensifier requires continued
operation during power mains interruptions, it is recommended that the EUT
image intensifier be powered from an uninterruptible power supply or a battery.
IEC 60601-1-2 Test Level
0 % UT; 250/300 cycle
Compliance Level
± 2 kV
Compliance Level
0 % UT; 0.5 cycle At 0°, 45°, 90°, 135°,
180°, 225°, 270° and 315°
0 % UT; 1 cycle and 70 % UT; 25/30
cycles Single phase: at 0°
Compliance Level
0 % UT; 250/300 cycle