Circulating Assistant Action

WARNING: DO NOT contaminate the aseptic
housing. The transfer shield guards the aseptic
housing from contamination during battery
installation.
4. Hold the non-sterile battery by the handle.
5. While aligning the battery's housing profile with
the transfer shield and chamber opening, place
the non-sterile battery into the chamber.
6. Lay the battery handle flat against its housing.
7. Remove the contaminated transfer shield and set
it aside.
6
WARNING: Sterilize the contaminated
transfer shield before using it again. See the
care instructions manual supplied with the
handpiece.

Sterile Assistant Action

8. Close the aseptic housing lid and move the latch
to the lock symbol. Ensure the dot on the latch
aligns with the dot on the lock symbol.
WARNING: Ensure the dot on the latch aligns
with the dot at the lock symbol. Failure to
comply may allow the lid to open, exposing the
non-sterile battery and result in patient injury.
9. Slide the aseptic housing containing the installed
non-sterile battery firmly into the handpiece until it
snaps into place.
6126-120-700 Rev-AA
www.stryker.com