Reference Conditions: 23°C ± 5°C (30 to 50% RH) range. Add 25 ppm/°C from -10 to 18°C and 28 to 50°C to all accuracy
specifications. No external electrical or magnetic fields. Output current ≤150 mA for VT/PT and ≤ 50 mA for CT. Calibration cycle
is 1 year.
3.1. DTR

8510 SPECIFICATIONS

®
ELECTRICAL
Ratio Range (VT/PT)
Accuracy (VT/PT)
Ratio Range (CT)
Accuracy (CT)
Excitation Signal
Excitation Current Display
Excitation Frequency
Display
Measurement Method
Power Source
Battery Life
Battery Charger
Charging Time
Data Storage
Date/ Time
Communication
Software
MECHANICAL
Dimensions
Weight
Connection
Leads
Enclosure
Vibration
Shock
Drop
Index of Protection
3. SPECIFICATIONS
Autoranging: 0.8000 to 8000:1
Ratio Range
0.8000 to 9.9999
10.000 to 999.99
1000.0 to 4999.9
5000.0 to 8000.0
Autoranging: 0.8000 to 1000.0
Ratio Range
0.8000 to 1000.0
VT/PT Mode: 32 Vrms max
CT Mode: Auto Level 0 to 1 A, 0.1 to 4.5 Vrms
Accuracy: ± (2% of Reading + 2 mA)
Dual line alpha-numeric LCD, 16x2 characters with contrast
adjustment and backlight control. Day/night visible.
In accordance with IEEE Std C57.12.90™
Two 12V, 5x2, 1650 mAH NiMH rechargeable battery packs
Up to 10 hrs of continuous operation. Low battery indication.
Universal input (90 to 264 Vrms input) smart recharger
99 objects with 99 tests each
Battery-backed, Real-time clock
USB. 2.0 compliant, optically isolated, 115.2 kB
DataView
272 x 248 x 130 mm (10.70 x 9.76 x 5.12")
4.6 m (15 ft+) H & X shielded with large color-coded
industrial alligator clips in carrying bag
Heavy duty Polypropylene case, UL 94 V0
IEC 68-2-6 (1.5 mm to 55 Hz)
IP 40 (Instrument lid open) per EN 60529
IP 53 (Instrument lid closed) per EN 60529
48
Accuracy (% of Reading)
Accuracy (% of Reading)
Range: 0 to 1000 mA
70 Hz
< 4 hrs
analysis software included
®
3.7kg (8.1 lbs)
XLR connectors
IEC 68-2-27 (30 G)
IEC 68-2-32 (1 m)
± 0.2%
± 0.1%
± 0.2%
± 0.25%
± 0.5%

Publicidad

Tabla de contenido
loading