Caution: Not cleaning the device as instructed can result
in contamination of the device.
EN
1. Use the cleaning wipe provided with each Bloom Test
to wipe the Bloom Lab surface after each use.

8. Store-use operating steps

For the use of in-store Bloom Labs (device usable by mul-
tiple users), please follow the instructions below.
The following steps should be carried out by a store employee:
Step 1: Set up Bloom Lab
Caution: Do not use the Bloom Lab if the packaging or
the device is obviously damaged. If you do so, the correct
functioning cannot be guaranteed.
Caution: do not use the Bloom Lab when it is exposed
to environmental conditions such as magnetic or
electric fi elds, external electrical infl uences, electrostatic
discharge, or thermal ignition sources.
Caution: Do not use the Bloom Lab with power supplies
other than those provided by Bloom Diagnostics in order
to avoid electronic malfunction of the device or electric shock.
Caution: Using the Bloom System in a dry environment,
especially if synthetic materials are present (synthetic
clothing, carpets, etc.) may cause damaging electrostatic
discharges that may produce erroneous results.
1. Unpack the contents of the Bloom Lab packaging.
2. Plug in the Bloom Lab using the provided power
supply. You will see the white LED ring on the front
lighting up and turning clockwise (see section
13.1.). This will take less than 1 minute.
3. Once the Bloom Lab is ready this will be indicated
by the white LED ring lighting up permanently (for
pictures see section 13.1.).
Step 2: Setting up an administrator device
8
1. Open the Bloom App on the smartphone you want
to use as the administrator device.
2. Log into the Bloom App.
3. Click "Add Test".
4. Tap the upper right corner of your smartphone to
the Bloom Lab. When pairing with the Bloom Lab
is successful, you will see a white heartbeat pattern
on the LED ring (see section 13.1.).
5. The first smartphone connected to the Bloom Lab
is set as the administrator device. The administrator
device can be used to update the Bloom Lab.
Optional step: Changing the administrator device
1. To reset the administrator device please press the reset
button on the bottom of the device (Visual N°10).
2. After resetting the device, the first smartphone
connected with the Bloom App will be defined as
the administrator device.
Step 3: Renaming the Bloom Lab
1. After connecting the Bloom Lab to the Bloom
App, please rename the Bloom Lab to give it an
individual identifier.
To use the Bloom Lab in-store users should now perform
steps 2–4 of section 7.
Step 4: Clean the Bloom Lab
Caution: Please use protective equipment, such as
gloves, when cleaning to avoid contamination.
Caution: Not cleaning the device as instructed can
result in contamination of the device.
1. Use the cleaning wipe provided with each Bloom
Test to thoroughly wipe all surfaces of the Bloom
Lab after each use.
2. Check gaps of the Bloom Lab casing for stains and
clean again when needed.
Step 5: Switch off your Bloom Lab
1. Unplug the power supply from the device after the
completion of the last measurement.
4-3-6-2 Bloom Lab IFU, 09.04.2020, V1.0