Stryker PINPOINT PC9000 Manual Del Operador página 100

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 984
WARNING:
The pre-cleaning steps described in preparing the PINPOINT camera should
always be performed prior to cleaning the PINPOINT camera.
WARNING:
Strictly adhere to the instructions provided by the cleaning agent
manufacturer to determine concentration and exposure time.
CAUTION:
Excessive concentrations of cleaning solution or excessive soaking time can
permanently damage the PINPOINT camera.
CAUTION:
Do not use the following substances as cleaning agents or disinfectants:
CAUTION:
Do not scratch contaminants off the camera using hard instruments as this can
damage the lens, cable, or buttons.
After preparation and following the General Practice guidelines, manually clean the
PINPOINT camera according to the following detailed manual cleaning instructions.
94
Organic, mineral or oxidizing acids (minimum acceptable pH 5.5)
Strong alkalis (maximum acceptable pH 11; neutral/enzymatic or slightly
alkaline detergent recommended)
Organic solvents (for example., alcohols, ethers, ketones, benzenes)
Oxidizing agents (for example, hydrogen peroxide)
Halogens (for example, chlorine, iodine, bromine)
Aromatic/halogenated hydrocarbons
Oils
Use freshly prepared solutions only.
Use sterile or low-bacteria (up to 10 bacteria/ml) and low-endotoxin (maximum
0.25 EU/ml) water (e.g., purified or highly purified water).
Use filtered air for drying.
Corrosion on the product can occur from using chloride-containing water,
therefore, fully demineralized water should be used for the final rinse
procedure
The cleaning agents used must be dispensed according to the prescription of
the manufacturer.
4-0002428 Rev. J

Publicidad

Tabla de contenido
loading