To sterilize laparoscopes and light guide cables using the NOVADAQ Sterilization
Tray (PC9017):
WARNING:
Using the incorrect filters will result in the product not being properly
sterilized and may lead to the failure of the sterile barrier.
All four filters must be replaced with new filters prior to sterilization.
Figure 50: Installing filters in the sterilization tray bottom
88
Insert four new filters, provided in the sterilization accessories kit (PP9076 or
PP9066), two each in both the sterilization tray bottom and lid as shown in
the following figures. The tray must be clean and dry prior to use.
4-0002428 Rev. J