The video image appears
too red or blue.
Image quality is poor due
to poor color or brightness.
Other equipment in the
vicinity of PINPOINT
seems to malfunction
when PINPOINT is
powered on, but works
normally when PINPOINT
is powered off.
PINPOINT Endoscopic Fluorescence Imaging System Operator's Manual
The Red or Blue
Saturation value is
too large
Image quality
settings have been
altered significantly
Radio frequency
interference may be
occurring between
PINPOINT and the
malfunctioning
equipment.
1.
Check the Image Quality value for the
Red or Blue Saturation using the option
menu.
2.
If the Red value is greater than 768 or
the Blue value is greater than 1024,
decrease the value.
Reset the Image Quality settings by selecting the
Default Profile from the user options menu.
1.
Check to see if any of the following
resolve the problem:
Reorient or relocate PINPOINT or
the equipment receiving the
interference.
Increase the separation between
PINPOINT and the equipment
receiving the interference.
Connect PINPOINT to a power
outlet on a different circuit from that
to which the other equipment is
connected.
2.
If PINPOINT becomes unresponsive,
and does not resume normal function
after turning off and back on, stop using
PINPOINT and contact a qualified
NOVADAQ service representative.