Table 17: Light Guide Cable Specifications
Optical
Physical
Environment
Sterilization
Table 18: Equipment Classification
Type of protection against electric shock
Degree of protection against electric shocks
Degree of protection against moisture
Laser class
Radio frequency emissions
Harmonic emissions
124
Transmission spectrum
Fiber diameter
Length
Weight/length
Operating temperature
Storage temperature
Refer to Table 10: Validated Sterilization Methods for NOVADAQ Laparoscopes
and Light Guide Cables
4-0002428 Rev. J
Visible + NIR
4.9 mm
3 m
125 g/m
+10 to +40°C
-40 to +70°C
Class I
CF-type
Ordinary
Class 3R
Complies with 21CFR 1040.10 and 1040.11 except for
deviations pursuant to Laser Notice No. 50, dated June
24, 2007.
Group 1, Class A
Class A
as per IEC 60601-1
as per IEC 60825-1
as per CISPR 11
as per IEC 61000-3-2