Descargar Imprimir esta página

Maintenance And Care; Device Disposal - Integra MAYFIELD A2101 Manual De Instrucciones

Unidad ultrabase

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 36
EN – English
8. Inspect the equipment to make sure there is no visible organic
debris or residue from cleaning agent.
Repeat process if any soil is detected.
Drying
• Products should be dry at this point. If wetness or excess liquid
is detected, dry with a soft clean towel.
• Medically compressed air can be used if needed.
Optional Automatic Wash / Disinfect
CAUTIONS
• Alkaline and highly acidic detergents and solutions cause
damage to the devices.
• Channels and crevices found on these devices require particular
attention during cleaning.
• Pay special attention to the water quality used throughout
reprocessing. Hard water can damage the surface of the
equipment. Avoid using hard water. Instead use purified water
unless otherwise specified.
Limitations on Reprocessing
• Repeated processing has negative effects on these devices and
is not recommended for routine use.
• It is important to have Integra NeuroSpecialists perform routine
inspections (twice yearly is recommended).
• See contact information.
INSTRUCTIONS
Cleaning - Automated Equipment: Neutral pH Detergent
Method
1. Prepare equipment for cleaning by disassembling removable parts
and loosening connections.
2. Rinse equipment in warm water before placing into washer.
3. Load device into the washer and place small parts in container
or tray inside the washer unit in order to avoid losing small
components. NOTE – Load devices carefully into washer in order to
avoid collision.
4. Follow the instructions listed below and set washer machine to
these exact parameters:
Phase
Time
Water
(Min.)
Temperature
Pre-wash 1
4:00
Cold water
Enzyme
4:00
Hot water
Wash
Wash 1
10:00
60.0ºC
(140º F)
Rinse 1
0:30
Hot water
Thermal
2:00
82.2ºC
Rinse**
(180º F)
Detergent and
Concentration
N/A
Neutral pH enzymatic (e.g.
Endozime® AW Triple Plus
with APA, Ratio 1:128)
Neutral pH detergent
(e.g. Renu-Klenz™ , Steris,
Ratio 1:256)
N/A
N/A
** Optional phase for disinfection of components – minimum water
temperature as indicated or per manufacturer specifications for the
thermal rinse cycle.
NOTE – Any deviation from this guideline could result in damage to
the equipment as well as improper cleaning results.
Rinse with purified water. Do not perform if parameters cannot
be achieved.
5. Remove from washer and dry completely if needed.
6. Inspect equipment to make sure there is no visible organic debris
or residue from cleaning agent.
Repeat process if any visible soil is detected.
The life expectancy of the MAYFIELD products is expected to be as
long as 7 years.

Maintenance and Care

To ensure proper function and to extend the life and performance of
the equipment, Integra LifeSciences recommends the following:
Recommended Action
Return the device to the Integra
LifeSciences Repairs department for
detailed inspection and servicing.
Request that Integra NeuroSpecialists
perform routine inspections of the
device.
In the absence of proper care and servicing of the device, negative
effects may be seen after repeated processing over time which may
lead to reduced performance.
Contact information: See the Service and Repair section for contact
information on how to return your device for periodic servicing and
to request periodic inspections.
See Inspection and /or Service notes section for routine checks to be
performed on the device.
NOTE – Any serious incident that has occurred in relation to the
device for the user and/or the patient should be reported to the
manufacturer and the competent authority of the member state in
which the user and/or patient is established.

Device Disposal

NOTE – Follow hospital procedures for disposal of this device.
6
Recommended Frequency
Once / year
Twice / year

Publicidad

loading

Este manual también es adecuado para:

Mayfield a2102