INDICATIONS FOR USE
The reusable AEROECLIPSE* XL Breath Actuated Nebulizer (BAN) is a single patient, reusable device when
used in a home environment and multi-patient when cleaned according to Hospital instructions. The nebulizer
is intended to administer prescribed aerosolized medication to spontaneously breathing patients under the care
or treatment of a licensed healthcare professional. The device may be used with an Ombra* Compressor or an
air source capable of delivering a flow of 3.5 to 5.0 litres per minute (Lpm) under a nominal operating pressure
of 0.92 to 1.79 bar (13.3 to 26 pounds per square inch [psi]). The intended environments for use include the
home, hospitals and clinics.
IMPORTANT SAFETY INFORMATION
The following statements will be used to draw attention to potential use errors that could cause harm.
NOTE - Indicates information to which the user should pay special attention.
CAUTION - Information for preventing damage to the product or causing personal injury.
WARNING - Important safety information for hazards that could cause serious injury or death.
BEFORE YOU BEGIN
1. This package should contain one reusable AEROECLIPSE* XL BAN with mouthpiece, one EZ Twist
TUBE
and one Owner's Manual (See Figure 1). If any parts are missing, please contact the authorized
distributor from whom you purchased this unit. To download an electronic copy of this Owner's Manual, visit
www.aeroeclipse.com
2. Prior to use, carefully examine the product for foreign objects, damaged or missing parts. Inspect the tubing
for fray, wear or loose particulate, and ensure nothing is blocking the air supply pathway of the nebulizer
cup. If any parts are missing or a defect is noticed, contact the authorized distributor where you originally
purchased your reusable AEROECLIPSE* XL BAN. Ensure you clean your nebulizer before first use and
between each treatment.
OWNER'S MANUAL
3. Read your
before starting a nebulizer treatment. If you have questions about the performance or usability of this
product, please contact your healthcare professional prior to use.
4. The nebulizer should be replaced after 6 months of use. Be sure to follow the use and cleaning instructions
as outlined in this Owner's Manual, and replace your nebulizer within the timeframe indicated to ensure it
functions correctly. Failure to follow the appropriate cleaning instructions may have an adverse affect on
device performance.
5. It is recommended that you have an extra nebulizer on
hand in the event your nebulizer becomes damaged, lost or
otherwise unusable.
SETTING UP YOUR NEBULIZER
For nebulizer part descriptions see Figure 1.
CAUTION: The reusable AEROECLIPSE* XL BAN can
deliver aerosolized medication in either
CONTINUOUS
or
Mode. In
medication is delivered on inspiration only and may result in an
increase of the dosage to the lungs. Clinicians should take this
information into consideration when determining dosage.
1. Prior to use, inhale and exhale through the Mouthpiece
to verify the movement of the
BUTTON
located on the top of the device (Fig. 1). The
button should move up and down freely. If not, move the
MODE SELECTOR
EN
2
to be sure you understand all instructions for use, cautions and warnings
BREATH ACTUATED
BREATH ACTUATED
GREEN FEEDBACK
BREATH ACTUATED
to the
FIGURE 2
Breath Actuated Mode
Mode,
position
Continuous Mode