Descargar Imprimir esta página

Thermo Scientific Sensititre AIM Instrucciones De Uso página 50

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 319
Table 17. Haemophilus influenzae reproducibility of AIM system vs. AutoInoculator™ system
(Assuming off scale result
is within one well from the
Vizion™ system
*Off-scale isolates were used due to unavailability or insufficient number of isolates with on-scale MIC values for some
drug/organism combinations.
Table 18. Yeast reproducibility of AIM system vs. AutoInoculator™ system
(Assuming off scale
result is within one well
Mirror Read
*Off-scale isolates were used due to unavailability or insufficient number of isolates with on-scale MIC values for some
drug/organism combinations.
50
% Reproducibility (AIM system vs. AutoInoculator™ system)
Best case
mode)
99.7
% Reproducibility (AIM system vs. AutoInoculator™ system)
Best case
from the mode)
99.7
On scale (On scale
values only)
99.8
On scale (On scale
values only)
99.7
Sensititre AIM™ Instructions for Use
*Worst case
Assuming off scale result is
greater than one well from the
mode)
99.4
*Worst case
(Assuming off scale result is
greater than one well from the
mode)
83.7

Publicidad

loading

Este manual también es adecuado para:

V3020