Operating instruction – method and instructions for use
2
Safety Notes
WARNING
The safety of persons, operation and processing quality with this laboratory apparatus depends primarily on the actions
of the persons who work with the apparatus.
Read the operating instructions carefully before using the apparatus and observe the directions for avoiding errors and resulting
damage, particularly damage to health.
In addition to the directions in the operating instructions, the applicable national laws, regulations and guidelines must also
be observed for installation and operation.
Maximum lift speed 50 mm / sec. Flip the switch upwards only when the fill cylinder is inserted.
Caution: danger of injury! Before working on the pressure module (pressure plunger) disconnect the apparatus from the
compressed-air supply!
NOTE
Keep the operating instructions in an accessible position near the apparatus!
3
Designated Use
The Palajet is a pneumatic injection apparatus for the production of dentures with PalaXpress
cold-curing polymer or heat-curing polymers such as Paladon
Area of application: dental technology, such as laboratories, schools, universities.
The manufacturer does not accept liability resulting from use that is not in accordance with the instructions for use
or unauthorised changes.
The apparatus is designed for continuous operation. Personal protective equipment, such as safety gloves, safety glasses etc.,
must be used at all times for protection from contamination with the processed materials.
4
Contents of delivery and accessories
Palajet contents of delivery
■
Injection cylinder including cover
■
1 Palajet Duoflask
(incl. allen-wrench (13 mm), opening key for flasks, distance holder, placeholder)
■
12 injection cylinder cartridges
■
12 cover cartridges
■
50 sealing disks
■
Measurement cup
■
Mixing cup
■
Mixing spatula
■
∅
Channel wax (33 strips
■
Timer
■
Flask filling basket
Palajet accessories
■
Palajet singleflask
(incl. allen-wrench (13 mm), opening key for flasks, distance holder)
■
Palajet Duoflask
(incl. allen-wrench (13 mm), opening key for flasks, distance holder, placeholder)
■
Injection cylinder including cover
■
Timer
■
Flask filling basket
5
Equipment Characteristics
■
The Palajet is a pneumatic injection apparatus for injecting denture material into denture flasks.
■
The apparatus is factory-set to a working pressure of 4 bar, which can be controlled with the pressure meter
(see paragraph 10 Sources of error).
■
The pneumatic feed pressure of max. 10 bar must not be exceeded.
■
The plunger speed during injection is 8 – 10 mm / sec.
■
The Palajet is designed and manufactured according to the most common safety regulations and guidelines.
Injection process for manufacturing dentures
®
∅
7 mm and 30 strips
3 mm)
65.
ultra and PalaXpress
®
®
Timer
GB
- 3 -