INSTRUCTIONS FOR USE
Setup:
1. Place the two rail clamps on the surgical table's back
section rail approximately 12", 30cm, apart on the
patient's affected side.
2. Hold the table vertically above the rail clamps with
the frame braces outboard of the surgical table and
the Mounting Blades pointing toward the open slots
of the rail clamps.
3. Lower the blades into the clamps & tighten the rail
clamp locking knobs.
4. Fold the table down making sure that the frame
brace ends engage the brace openings in the
mounting blades.
5. To adjust height, loosen the rail clamp knobs. Grasp
the table top adjacent to the rail clamps and raise or
lower the table to the desired height. Securely tighten
the rail clamp knobs.
6. O-AHTR and O-AHTH Only. With the Table set at the
surgeons working height, loosen Leg Lock Handle at
the ball socket near end of table. Swing the leg down
and slide ball socket to position of least interference with surgeon, C-arm, etc.
Loosen Leg Extension Knob and adjust leg length so the foot is securely against
the floor. Firmly lock both handles to stabilize table end.
Note: Support leg may be folded up out of the way for C-arm access only. Care must
be taken not to apply any additional loads beyond the patient's arm weight while
the support leg is not in contact with the floor. (O-AHMPT do not have a leg).
Device controls and indicators:
Controls and indicators of this device are described within the Setup instructions.
Storage, Handling and Removal Instructions:
3.4.1
Storage and Handling:
The product should be stored in a clean and safe environment to prevent product
damage. See storage Specifications under Product Specification section.
Document Number: 80028133
Version C
Page 13
Issue Date: 27 MAR 2020
Ref Blank Template: 80025117 Ver. F