Icon
Text
Preheat
P Ramp
Ramp
Time
Auto Start
Auto Stop
iBreeze CPAP System User Manual
Figure3 Patient Comfort Setting Interface
Description
Turn Preheat function on or off. When the Preheat is on, the
humidifier starts preheating in the standby mode and the
maximum preheating time is 30 minutes. In the therapy mode,
the preheating stops.
Option: ON / OFF
Note:
1. If the humidifier water level below the limit, Preheat function
will be turned off automatically.
2. If humidifier function is turned off, Preheat function will be
disable.
3. Preheat function is not available in the iBreeze 20C.
Set the starting pressure of Ramp.
Setting Range: 3 cmH
increments.
Default: 4 cmH
O
2
Set the increase time from Ramp pressure to the setting
therapy pressure.
Option: 0-60 mins, 5 mins increments
Default: 15 mins
Note: if Ramp Time sets to 0 min, Ramp function will be off.
Turn Auto Start function on or off.
When the Auto Start function is on, the system will start
therapy automatically if a breath with mask is detected.
Option: On / OFF
Turn Auto Stop function on or off.
When the Auto Stop function is on, if the mask is removed
more than 5 seconds, the therapy mode will stop.
Option: On / OFF
Default: OFF
O-Setting pressure, 0.5 cmH
2
Default: ON
Default: ON
16
O
2