Resvent iBreeze CPAP System Manual Del Usuario página 28

Publicidad

Idiomas disponibles

Idiomas disponibles

Alert
Low Respiratory
Rate (RR).
Apnea.
High Leakage
Volume.
High Inspiration
Pressure.
Low Expiratory
Pressure.
Mask vent holes
blocked.
Low Tidal Volume
(Vt).
System Drying
Low battery
charge, reports
may not be saved.
RTC clock failure.
Reports may not be
saved.
Note: Some alert messages depend on the device model.
iBreeze CPAP System User Manual
Type
Possible Cause
threshold value.
Alert 0
None.
An apnea is detected
Alert 1
and exceeds the
setting duration.
1. Inappropriate
connection of mask
Alert 1
and respiration tube.
2. The water tank is not
plugged in.
During breath cycle
monitored pressure is
Alert 0
higher than the set
pressure threshold.
During breath cycle
monitored pressure is
Alert 0
lower than the set
pressure threshold.
Mask vent holes is
Alert 1
blocked.
During breath cycle,
Alert 0
Average Vt less than
0.6*Setting Value.
The device generates
a small flow to dry the
system when it is the
Notification
standby mode after 30
minutes usage with
humidification.
Low button battery
Notification
power.
Alert1
RTC clock failure.
25
Action
parameter setting.
Check the respiration
tube or adjust therapy
parameter setting.
Check the respiration
tube or adjust therapy
parameter setting.
Check the connection
of the mask or
respiration tube, and
the connection of the
water tank.
Check the respiration
tube or adjust therapy
parameter setting.
Check the respiration
tube or adjust therapy
parameter setting.
Check the mask.
Check the respiration
tube or adjust therapy
parameter setting.
No action.
Please contact the
supplier if you want
reports be saved.
Please contact the
supplier if you want
reports be saved.