15 Specification
Physical
Dimension (L*W*H): 238*178*128 mm
Weight: Approximately 1.60 kg
Operating Environmental
Temperature: 5℃~35℃
Relative Humidity: 10%~95% (non-condensing)
Atmospheric Pressure: 70 kPa~106 kPa
Storage Environmental
Temperature: -25℃~70℃
Relative Humidity: 5%~95% (non-condensing)
Atmospheric Pressure: 70 kPa~106 kPa
Noise Value
Sound pressure level Uncertainty
28 dB(A)
Standards compliance
IEC 60601-1:2005 Medical electrical equipment - Part 1: General requirements
for basic safety and essential performance.
IEC 60601-1-2:2014 Medical electrical equipment - part 1-2: general
requirements for basic safety and essential performance - collateral standard:
electromagnetic disturbances - requirements and tests.
ISO 80601-2-70:2015 Medical electrical equipment - Part 2-70: Particular
requirements for basic safety and essential performance of Sleep apnea
breathing therapy equipment.
ISO 8185:2009 Respiratory tract humidifiers for medical use - Particular
requirements for respiratory humidification systems.
IEC 60601-1-11:2015 Medical electrical equipment - Part 1-11: General
requirements for basic safety and essential performance –Collateral Standard:
Requirements for medical electrical equipment and medical electrical systems
used in the home healthcare environment.
Electrical
Specified power supply: Resvent Medical Technology Co., Ltd.
Model: DA60-220S24-W1
AC Input: 100-240V~1.18-0.56A 50/60Hz 2A Max
DC Output: 24 V⎓2.5 A
Specified power supply: SHANGHAI BIAOJUN ELECTRONICS
TECHNOLOGY Co., Ltd.
Model: JE01-40-006HM
AC Input: 100-240V~50/60Hz 2A Max
DC Output: 24V⎓2.5 A
iBreeze CPAP System User Manual
Sound power level
2 dB(A)
36 dB(A)
31
Uncertainty
2 dB(A)