10 Device Alert
There are 4 types of alerts described here::
Alert 0: Gray bottom white tips, no light, no sound, always display, disappear until
the prompt condition not met.
Alert 1: Orange bottom white tips, blue-ray flashing, with sound, always display,
disappear until the prompt condition not met.
Alert 2: Red bottom white tips, blue-ray flashing, with sound, exclusive alarm until
the user opens the alert to confirm or press the Therapy On/Off key.
Notification: White reminder message, no light, no sound, after the
corresponding process, the message automatically disappears.
Alert Summary Table: The following table summarizes the alerts.
Alert
Input voltage is
abnormal, please
check!
SD card write/read
underway, do not
remove the SD
card, do not cut off
the power.
SD Card Removed.
SD Card Full,
Please replace the
SD Card.
SD card can't be
written, please
unlock and insert
again.
SD card error,
please remove and
insert again.
iBreeze CPAP System User Manual
Type
Possible Cause
Power Adapter with the
wrong type, resulting in
Alert 2
voltage is too high or
too low
1. Insert SD card
during data
synchronization;
Notification
2. Input the
configuration on the
SD card.
1. No SD card in the
device.
Notification
2. SD card has been
removed.
In Standby mode, SD
Notification
card storage space
only 200M
The SD card is
Notification
read-only and can't be
written.
SD card failure, may
be: 1.SD card can't
Notification
read and write.
2.SD card read and
write data errors.
23
Action
Use the Power
Adapter provided by
Resvent.
No action.
Reinsert functional
SD card.
Replace the SD card
or clean the data after
export the data in the
SD card.
Remove the SD card,
unlock and insert it
again.
Remove SD Card,
Reinsert or replace
with a new card.