Safety Specifications
Class II
Type BF
Ingress Protection: IP22
Air filter
Air filter: Filter Efficiency: >75% (7 micron dust)
Pressure
Setting Range: 4-20 cmH
Max Single Fault Steady Pressure: 40 cmH
Static Pressure Accuracy: ±0.5 cmH
Dynamic Pressure Accuracy: ±1cmH
Pressure monitoring Accuracy: ± (2%*full scale reading + 4%*actual reading)
Flow
Measured pressure at the
patient connection port
19mm
(cmH
breathing
tubing
Average flow at the patient
connection port (L/min)
Measured pressure at the
patient connection port
15mm
(cmH
breathing
tubing
Average flow at the patient
connection port (L/min)
Humidifier
Water capacity: 290 ml (MAX Water Level)
Humidity: >10 mg/L BTPS (Within the set pressure range)
Wrist Pulse Oximeter:
Model: MD300W314B4
SpO
SpO
Pulse Rate Range: 30~250 bpm
Pulse Rate Accuracy: 30~99 bpm: ±2 bpm; 100~250 bpm: ±2%
Note: refer the user manual (version: Ver1.0w314B4) for more details.
iBreeze CPAP System User Manual
O
2
O)
2
O)
2
Range: 70%~100%
2
Accuracy: 70%~100%: ±2%; <70%: unspecified
2
O
2
O (Expanded uncertainty:1.87%)
2
O (Expanded uncertainty:2.55%)
2
Test pressure (cmH
4
8
2.9
7.1
154.3
146.4
2.9
6.9
141.4
132.7
32
O)
2
12
16
20
11.0
15.1
19.1
135.7
125.3
114.1
10.8
14.8
18.7
124.2
114.3
85.4