Form A-2 Guidance and Declaration - Electromagnetic Immunity
Immunity test
Electrostatic
discharge ( ESD )
IEC61000-4-2
Electrical fast
transient/burst
IEC61000-4-4
Surge
IEC61000-4-5
Creux de
tension, court
interruptions et
variations de
tension
sur l'alimentation
lignes d'entrée
IEC61000-4-11
Power frequency
( 50/60Hz )
magnetic field
IEC61000-4-8
Conduced RF
IEC61000-4-6
IEC60601test
level
Contact : ±8kV
Air : ±15 kV
Power supply
lines : ±2 kV
input/output
lines : ±1 kV
line(s) to line(s) :
±1 kV
line(s) to earth :
±2 kV
0%, 70%, 0% of
UT
50Hz,60Hz
30A/m
150KHz to
80MHz 3Vrms
ISM and amateur
radio bands
Compliance
level
Contact : ±8 kV
Air : ±15V
Power supply
lines : ±2 kV
input/output
lines : ±1 kV
line(s) to
line(s) : ±1 kV
line(s) to earth :
±2 kV
0% for 0.5 cycle
0% for 1 cycle
70% for 25
cycles
0% for 250
cycles
50Hz:30A/m
60Hz:30A/m
3Vrms
6Vrms (in ISM
and amateur
36
Electromagnetic
environment
guidance
Les planchers doivent
être en bois, en béton
ou en carreaux de
céramique. Si les sols
sont recouverts de
matière synthétique, le
l'humidité relative
devrait être d'au moins
30%.
La qualité de
l'alimentation secteur
doit être celle d'un
environnement
commercial ou
hospitalier typique.
La qualité de
l'alimentation secteur
doit être celle d'un
environnement
commercial ou
hospitalier typique.
La qualité de
l'alimentation secteur
doit être celle d'un
environnement
commercial ou
hospitalier typique.
La qualité de
l'alimentation secteur
doit être celle d'un
environnement
commercial ou
hospitalier typique.
Les équipements de
communication RF
portables et mobiles
ne doivent pas être
-