Alert
Non-original
card,continue using
please format.
Software update
underway, Do not
cut off the power!
System error code:
XXXX Please try to
restart, please
contact the supplier
if repeat.
The respiration
tube expired,
please replace.
The water tank
expired, please
replace.
The filter expired,
please replace.
The mask expired,
please replace.
It is time for device
maintenance,
Please contact the
service provider for
device
maintenance.
Low Minute
Volume (MV).
High Respiratory
Rate (RR).
iBreeze CPAP System User Manual
Type
Possible Cause
Insert other device' SD
Notification
card in non-service
mode.
Notification
Software update.
1. Pressure sensor
failure, flow sensor
failure, blower failure in
Alert 2
therapy state; 2. Power
board short-circuit 3.
Humidifier heating
circuit short-circuit
In standby state, the
usage time exceed the
Notification
respiration tube setting
service life.
In standby state, the
usage time exceed the
Notification
water tank setting
service life.
In standby state, the
usage time exceed the
Notification
filter setting service
life.
In standby state, the
usage time exceed the
Notification
mask setting service
life.
In standby state, the
usage time exceed the
Notification
setting device
maintenance time.
Minute volume less
Alert 0
than the setting
threshold value.
Respiratory Rate
Alert 0
exceed the setting
24
Action
Format the SD card in
your computer and
then insert it to
device.
No action.
Please try to restart,
please contact the
supplier if repeat.
Click "Confirm",
replace the respiration
tube.
Click "Confirm",
replace the water
tank.
Click "Confirm",
replace the filter
Click "Confirm",
replace the mask
Click "Confirm",
contact the service
provider for device
maintenance.
Check the respiration
tube or adjust therapy
parameter setting.
Check the respiration
tube or adjust therapy