Problem
mask seems too low
(it feels like I am not
getting enough air).
I have stopped
therapy, but the
device is still blowing
air.
The device's display
is erratic (crash blank
or blue screen).
My water tank is
leaking.
Key exception
(non-responsive or
insensitive).
The knob is
insensitive.
The touchscreen is
not working.
iBreeze CPAP System User Manual
Possible Cause
progress.
Device is cooling
down.
The device has
been dropped or
mishandled.
Water tank may
not be assembled
correctly.
Water tank may
be damaged or
cracked.
Program crashes
or key
misalignment.
Encoder is
damaged
The touchscreen
is damaged
27
Action
Ramp Time off.
Device blows a small amount of air in
order to avoid condensation in the air
tubing. It will stop automatically after a
few minutes.
Unplug the device. Reapply power to
the device. If the problem continues,
contact your home care provider.
Check for damage and reassemble
the water tank correctly. Contact your
care provider for a replacement.
Unplug the device. Reapply power to
the device. If the problem continues,
contact your home care provider.
Unplug the device. Reapply power to
the device. If the problem continues,
contact your home care provider.
Unplug the device. Reapply power to
the device. If the problem continues,
contact your home care provider.