Application
3.2
Operation
Disinfect contaminated surfaces of the curing light as well as light guides and anti-glare cones before each use.
Additionally, the light guide can be sterilized using the autoclaves intended for this purpose (see chapter Maintenance
and Cleaning). Furthermore, make sure that the emitted light intensity permits adequate polymerization. For that
purpose, inspect the light guide for contamination and damage and check the light intensity at regular intervals.
Selecting the curing program and curing time
Bluephase N G4 is equipped with 4 selectable curing times and 3 curing programs for different indications.
Use the Time/Program selection button to set the desired curing time and/or the light intensity.
T
(Turbo Program), 2,000 mW/cm
*:
2
The curing time is set to 5 seconds in the Turbo program and cannot be changed.
H
(High Power Program), 1,200 mW/cm
*:
2
The following curing times can be selected in the High Power program: 10, 15 or 20 seconds.
PRE (PreCure Program)*:
The PreCure program is used to tack cure light-curing, adhesive luting composites in order to facilitate the removal
of excess material. If other luting composites are used, the distance from the light guide to the composite must be
increased or several curing cycles must be conducted. The curing time of the PreCure program is preset to 2 seconds
and cannot be altered.
* Rounded values. See Chapter 7 Product Specifications
The PreCure program must not be used for conventional light-curing!
10