3 Application
3.1
Start-Up
Check the delivery for completeness and any possible transportation damage (see List of Parts).
If parts are damaged or missing, contact your Ivoclar Vivadent representative.
Charging base
Before you switch on the device, make sure that the voltage mentioned on the rating plate complies with your local
power supply. Connect the power cord with the power supply. Make sure that the power cord is easily accessible at
any time and can be easily disconnected from the power supply.
Handpiece
Unpack the handpiece from the packaging and detach the light guide by
pulling it out. Then clean the handpiece and light guide (see Maintenance
and Cleaning). After cleaning, reinsert the light guide.
For reasons of hygiene, we recommend using a disposable protective sleeve
for each patient (see Maintenance and Cleaning). Make sure to fit the
protective sleeve snugly to the light guide. Sleeves enclosed in the delivery
form may be used or sleeves can be purchased per country specific
regulations. Then, attach the anti-glare cone or anti-glare shield to the
light guide.
Battery
The battery must be fully charged before it is used for the first time! When
fully charged, the battery features a curing capacity of at least 20 minutes.
Slide the battery straight into the handpiece until you hear and feel it click
into place.
Gently place the handpiece in the corresponding rest in the charging base
without using any force. If a hygiene sleeve is used, please remove it before
you charge the battery. If possible, always use the light with a fully charged
battery. This will prolong the service life. We therefore recommend placing
the handpiece into the charging base after each patient. If the battery is
fully discharged, the charging time is 2 hours.
The battery is an expendable part that typically needs to be replaced after
approximately 2.5 years. See battery label for the age of the battery.
Battery: charging status
The current charging status is displayed on the handpiece as described on
page 6.
9