Application
Cure Memory function
The last settings used, together with the combination of curing program and curing time, are automatically saved.
Start/Stop
The light is switched on by means of the Start/Stop button. It is recommended that the emission window of the light
guide is placed directly on the material to be polymerized. Once the selected curing time has elapsed, the curing
program is automatically terminated. If desired, the light can be switched off before the set curing time has elapsed
by pressing the Start/Stop button again.
Acoustic signals
Acoustic signals can be heard for the following functions:
– Start (Stop)
– Every 10 seconds
– Curing time and program change
– Insert battery
Light intensity
The light intensity is maintained at a consistent level during operation. If the supplied 10 mm light guide is used, the
light intensity has been calibrated to 1,200 mW/cm
± 10% in the High Power program. If another light guide than the
2
one supplied is used, it directly influences the emitted light intensity. In the Turbo program, the light intensity is
calibrated to 2,000 mW/cm
+/– 10%.
2
In a light guide with parallel walls (10 mm), the diameter of the light entry and the light emission window is the same.
When using focussing light guides (e.g. Pin-Point light guide 6>2 mm), the diameter of the light entry is larger than
that of the light emission window. The incident blue light is therefore concentrated over a smaller area. In this way,
the emitted light intensity is increased. Pin-Point light guides are suitable for spot-on polymerization, e.g. to fix
veneers before the removal of excess. For complete polymerization, the light guide must be changed.
12