Descargar Imprimir esta página

Bowa TissueSeal PLUS Instrucciones De Uso página 58

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 139
2 Safety
2.2.1.
2.2.2.
14
TissueSeal / TissueSeal PLUS
HF device
Use only approved HF devices and programs (see
section section 8, page 40).
Observe the operating instructions of the HF device and
the general instructions for electrosurgical operations!
Improper use of HF current can lead to endogenous and
exogenous burns and explosions:
Avoid direct skin contact with HF cables.
Avoid contact with flammable gases and liquids.
HF cable
Improper use of HF cables can lead to patient injuries:
Never lay the HF cable on the patient's skin.
Connect the ligation instrument for coagulation before
switching on the HF generator.
When plugging or unplugging the HF cable, always
grasp the connector directly.
Use only HF cables that are in perfect condition. Never
use a defective HF cable.
The HF cable may cause interference to imagery on
monitors:
Never route the HF cable alongside a camera cable.
Do not lay the HF cable in loops.
Consult the operating manuals of the BOWA HF
generators for additional information on interference with
other devices.
MN031-443-S0-20110112-EN

Publicidad

loading

Este manual también es adecuado para:

Tissueseal