MN031-443-S0-20110112-EN
Suitable cleaning agents
Ensure that the selected cleaning agent system fulfils
the following requirements:
•
The cleaning agent is suitable for the ligation
instrument.
•
If thermal disinfection is not used, a suitable
disinfection agent with proven effectiveness
(e.g. DGHM or FDA approval or CE mark)
and compatible with the cleaning agents is
also used.
•
The chemicals that are used are compatible
with the instrument components (see
section 7.11, page 39).
•
Follow the manufacturer's instructions with
regard to the concentration and exposure
time of the cleaning and disinfection
products.
TissueSeal / TissueSeal PLUS
7 Preparation
33