5 Operating Instructions
38
3.
Laser radiation:
ON when the foot control or finger switch is actuated,
OFF when the laser is in standby mode.
4.
Selected laser power: in our example, 0.8 W peak power. Using the
"RIGHT ARROW" and "LEFT ARROW" key, you can adjust the emitted
power between 0.5 W and 7 W in 0.1 W increments.
WARNING
Setting excessively high power levels entails the risk of overheating the pa-
tient's soft or hard tissue. Setting the power to excessively low levels may re-
sult in reduced treatment efficacy.
5.
Frequency: Modulation frequency of the laser. The frequency range
extends from 1 to 10,000 Hz. If set to zero, CW (for continuous wave
mode) is displayed. Up to 100 Hz, the parameter value can be increased
in 1 Hz increments. Between 100 Hz and 1,000 Hz, the parameter value
can be increased in 100 Hz increments. Between 1,000 Hz and 10,000
Hz, the parameter value can be increased in 1,000 Hz increments. If you
press the "RIGHT ARROW" key for more than 3seconds, the value
increases at a faster rate.
6.
Treatment energy: The laser unit calculates the energy used during
treatment (in mJ) from the power values and the selected laser activation
time. If no activation time has been preset, the energy will not be
calculated.
WARNING
Following program selection, the SIROLaser is in the standby mode. Any ac-
tuation of the foot control or finger switch activates the laser unit.
When you actuate the foot control or finger switch, the message "LASER ON"
appears. At the same time, four yellow LEDs light up, the audible alarm
sounds and the laser starts emitting. When you release the foot control or fin-
ger switch to interrupt treatment, the laser is deactivated, but remains ready
for operation.
To change the power, press the "ENTER" key. As soon as the area has been
selected, you can change the values by pressing the "RIGHT ARROW" and
"LEFT ARROW" keys.
Laser irradiation (and hence treatment) can be terminated immediately by
pressing the red "LASER STOP" emergency button.
To quit the selected menu and save your changes, press Change for more
than one second. A message stating that therapy has been terminated
appears on the display.
The "LEFT ARROW" key returns you to the main menu.