10 Manufacturer's declaration on electromagnetic compatibility
Interference immunity
The SIROLaser is intended for use in the electromagnetic environment specified below. The user should ensure that it is
used in such an environment.
Immunity test
Electrostatic discharge (ESD)
EN 61000-4-2
Electrical fast
transient/burst
EN 61000-4-4
Overvoltage
EN 61000-4-5
Voltage drops, brief power
outages and voltage fluctua-
tions on power supply input
lines
EN 61000-4-11
Magnetic field of line fre-
quency
EN 61000-4-8
60 56 092 D 3485
D 3485.201.01.10.02 03.2008
EN 60601-1-2
Compliance level
Inspection
stage
± 6 kV contact dis-
± 6 kV contact dis-
charge
charge
± 8 kV contact dis-
± 8 kV contact dis-
charge
charge
± 2 kV for power
± 2 kV for power sup-
supply lines
ply lines
± 1 kV push-pull
± 1 kV push-pull
< 5% U
< 5% U
T
(> 95% drop in U
)
(> 95% drop in U
T
for 0.5 cycles
for 0.5 cycles
40% U
40% U
T
(60% drop in U
)
(60% drop in U
T
for 5 cycles
for 5 cycles
70% U
70% U
T
(30% drop in U
)
(30% drop in U
T
for 25 cycles
for 25 cycles
< 5% U
< 5% U
T
(> 95% drop in U
)
(> 95% drop in U
T
for 5 seconds
for 5 seconds
3 A/m
3 A/m
Electromagnetic environment
- Explanation
The floors should be made of wood or
concrete or finished with ceramic tiling.
If the floor has a synthetic covering, the
relative humidity should be at least
30 %.
The power supply should be of a type
that is customary in business environ-
ments or in hospitals.
The power supply should be of a type
that is customary in business environ-
ments or in hospitals.
The power supply should be of a type
T
that is customary in business environ-
)
T
ments or in hospitals.
If the user must be able to use the
SIROLaser unit even in cases where
the power supply is interrupted, we rec-
T
ommend operating the SIROLaser unit
with an uninterruptible power source or
)
T
a battery.
T
)
T
T
)
T
he magnetic fields of the power line fre-
quency should move within a range that
is normal for a typical commercially
zoned or hospital site.