5 Operating Instructions

30
5
Operating Instructions
CAUTION
Always protect the optical fiber and its connector with the protective caps pro-
vided for this purpose after every treatment (see Fig. 15, page 24).
5.1

Warning and error messages

After selecting the indication, press the "ENTER" key. The laser unit is now
ready for operation. The message "Wear protective goggles" appears on the
display and four green LEDs start flashing.
If this message is displayed, check whether the optical fiber is properly con-
nected. As soon as this is the case, this message will no longer be dis-
played.
If this message is displayed, you must select an optical fiber diameter with
the "RIGHT ARROW" key and then confirm your selection with "ENTER".
If this message appears, you must insert the specified optical fiber and con-
firm this action with "ENTER".
This message indicates that the INTERLOCK safety device has responded.
Either the door of the treatment room is not closed, or the INTERLOCK con-
nector is not properly seated in its socket.
If this message is displayed, allow the laser to cool down for a few minutes.
The temperature of the laser diode must be < 5°C or >90°C; otherwise the
temperature sensor is defective.
Adapt the SIROLaser to room temperature over a period of 2 hours.
If this message nevertheless still appears, contact your dealer or an authorized
Customer Service Department.