7 Cleaning and sterilization
44
Steam sterilizers are approved for sterilization that fulfill the requirements of
EN 13060 or at least use fractionated vacuum and are suitable for the steril-
ization of dental handpieces such as e.g. SIRONA DAC PROFESSIONAL.
7.4

Cleaning the laser unit and the finger switch

Use a dry, soft cloth to remove dust from the SIROLaser. More stubborn
spots can be removed with a damp cloth.
You can disinfect the SIROLaser using any of the products that are commonly
used to disinfect medical electrical equipment, e.g. MinutenSpray -classic,
®
CAVICIDE
or Caviwipe.
CAUTION
Spray disinfection may allow liquids to penetrate into the laser unit!
The Sirolaser may be disinfected only by wiping it.
Observe the instructions provided by the manufacturers of these disinfec-
tants.
Disinfect the finger switch by spray or wipe disinfection.
You can use for instance: MinutenSpray classic from Alpro; MinutenWipes
from Alpro. In the USA: Cavicide
Observe the instructions provided by the manufacturers of these disinfec-
tants.
CAUTION
Make sure that the optical fiber is not damaged when removing the finger
switch.
WARNING
For further safety measures, see Chapter 1.3.
CAUTION
The SIROLaser unit cannot be sterilized.
CAUTION
The handpieces and reusable tips must be sterilized after each treatment.
®
TM
and Caviwipes
.
D 3485.201.01.10.02 03.2008
60 56 092 D 3485