Lámpara de quirófano

Publicidad

Surgical light
Lámpara de quirófano
Appearance:
Check the following points:
General condition of arm covers.
Condition of lenses.
Condition of seals.
Suspension system clean.
Dismantling certain elements may affect operation
and safety.
Examples include:
• Servicing the electrical power supply,
• Servicing the suspension arm and balance sys-
tem,
• Servicing the optical system of cupolas equipped
with filters designed to eliminate radiation not vis-
ible to the patient. Surgical lights must never be
used without these filters.
Contact the authorised MAQUET after-sales service depart-
ment for this type of inspection.
5.3
Recycling
Discharge bulbs, like commercial fluorescent tubes,
contain a proportion of mercury and must be dis-
posed of as special waste.
Halogen bulbs can be disposed of as domestic
waste.
User's manual / Manual de utilización
Aspecto:
Compruebe los puntos siguientes:
Estado general de las tapas de los brazos
Estado de las lentes
Firmeza de las juntas
Limpieza de la suspensión
El desensamblaje de algunos subconjuntos puede
afectar al funcionamiento y la seguridad. Por ejem-
plo:
• si se interviene en la alimentación eléctrica.
• si se interviene en los brazos de suspensión y el
sistema de equilibrado
• si se interviene el sistema óptico de las cúpulas
provistas de filtros para eliminar las radiaciones no
visibles para el paciente. La unidad no debe utili-
zarse sin estos filtros bajo ninguna circunstancia.
Para este tipo de intervenciones, contacte con el servicio pos-
venta autorizado por MAQUET.
5.3
Reciclado
Las lámparas de descarga, como los tubos fluores-
centes contienen mercurio y, por ello, deben ser so-
metidas a un reciclado especial.
Las lámparas halógenas se pueden eliminar con la
basura doméstica.