(e.g. muscle activity, placement of the pressure sensor on
bony protrusions, shape and circumference of the extremity,
elasticity of the compression material) have a direct impact on
the pressure values displayed. This means that the measured
values are constantly fluctuating. They can therefore merely
be seen as a guide. Thus, measurement of compression using
the Juzo  Pressure  Monitor must never replace observation of
the patient (e.g. mention of pain, noticeably pale and / or cold
extremity).
Juzo  Pressure  Monitor contents
1 hard case
1 monitor and control unit
1 sensor unit (consisting of Luer-Lock connecting hose, pres-
sure sensor, positioning aid)
1 pack containing 50 disposable protective covers (more can
be ordered)
1 Instructions for use
Important information
Use of batteries
Do not use a different type of battery for this device than
indicated in the battery compartment.
Do not insert the battery in the wrong direction (poles on the
wrong side).
A flat battery must be immediately replaced with a new one.
Remove the battery if the device will not be used for three
months or more.
If battery acid makes contact with your skin, rinse it immediate-
ly with a lot of clean water. Make sure you avoid contact with
your eyes. In the case of contact, consult a doctor immediately!
General use
The device has been exclusively designed to measure the
pressure of compression rolls or adjustable compression
devices. Do not exert excessive pressure when inserting
the positioning aid under compression garments. The
Juzo  Pressure  Monitor is not suitable for the pressure measure-
ment of compression stockings.
The Juzo  Pressure  Monitor has been designed to be used by
specifically trained medical professionals only.
Compression measurement must never replace patient obser-
vation (e.g. mention of pain, pale and / or cold extremity).
Do not leave the device unattended within the reach of small
children.
Use the sensor unit on undamaged skin only.
Do not exert excessive pressure when inserting the positioning
aid under compression garments.
Do not take the control unit apart.
The Luer-Lock connecting hose, pressure sensor and position-
ing aid make up the sensor unit. Do not take them apart!
The Luer-Lock connector has been exclusively designed to
connect the control unit with the sensor unit.
The disposable protective covers have been exclusively
designed for the sensor unit. They protect the pressure sensor
and help with hygiene. Use a new protective cover for each
patient.
The Juzo  Pressure  Monitor has not been designed for long-
term patient measurements.
The device was calibrated before delivery and should not
require maintenance for its usage period of 5 years.
You are welcome to return waste equipment to us free of
charge. For more information, please contact our Juzo
Customer Service.
General warnings
Ensure that the sensor unit is not damaged by pointed objects
or those with sharp edges.
Any damage to the control or sensor unit (in particular the pres-
sure sensor) can have a direct impact on the functionality of the
device. Only ever use the device if it is in perfect condition.
Before disposing of the device or its accessories, read and
follow the section "Storage, cleaning and disposal".
Elements and positioning of the
Juzo  Pressure  Monitor and control unit
1. Monitor and control uni
2. Luer-Lock connecting hose
3. Positioning aid
4. Pressure sensor
5. Disposable protective cover
6. Correct placement of the pressure sensor for measurement
Preparing the device for use
A. Before first use, please insert an appropriate battery. Please
note the battery requirements in the battery compartment.
Then connect the Luer-Lock screw cap at the end of the air
hose with the control unit. Ensure that the connection has
been screwed on tightly and cannot become loose by itself.
B. Insert the sensor unit into a disposable protective cover.
Do not use the Juzo  Pressure  Monitor without a protective
cover. In addition to hygienic reasons, it also protects the
pressure sensor from possible snapping off or slipping.
Using the Juzo  Pressure  Monitor
C. Press the Power button
on. Initially, the display will show "888" while the device au-
tomatically resets. Wait about 3 – 4 seconds until the display
to turn the Juzo  Pressure  Monitor