Spare Part
Thickness: 4 pt
1.6. Spare Part Replacement Process
This section specifies a spare part replacement process which is in compliance with IEC 62353.
Read these Instructions for Use carefully before performing any replacements on the
Ambu® aBox™ 2.
The technician and personnel being responsible for the replacement process should fulfil
the requirements outlined in section 1.1.
Follow the process in Figure 2 to replace spare parts in the displaying unit.
Use the Trouble
Shooting guide in
the IFU
Chapter 2:
Clean and disinfect
the displaying unit
2. How To Clean and Disinfect the Displaying Unit
CAUTIONS
•
Cleaning and disinfection wipes shall be moist, but not dripping to ensure no
damage to internal electronics of the displaying unit.
•
If using wipes containing hypochlorite or citric acid, ensure that all residue is
completely removed. Wipes containing hypochlorite or citric acid may affect the
screen's antireflective coatings over time. You should limit the use of wipes
containing hypochlorite or citric acid to required cases only.
The displaying unit is a reusable medical device. According to the Spaulding classification,
the displaying unit is a non-critical device.
The displaying unit should be cleaned and disinfected before and after repair and replacement
procedures following one of the procedures below.
Any deviation from the instructions should be properly evaluated for effectiveness and
potential adverse consequences by the person responsible for cleaning and disinfection,
to ensure that the device continues to fulfil its intended purpose. Cleaning procedures should
Description
Ambu® aBox™ 2 – Visualization device
interface kit – Grey-Empty-Green
Contains:
One grey and one green visualization
device interface board (VDI), a front cover
with a colour ring (grey and green), a
plectrum tool, and two screws for the VDI.
Power cables – J (CH), K (DK), I (AUS/NZ)
Power cable – B (US, JP)
Power cables – G (UK), E/F (EU, not DK, CH)
Contact Ambu, if
you need further
assistance
Chapter 3:
Chapter 4:
Ensure you have
Replace the
the tools needed
relevant part(s) of
the displaying unit
Follow the standard operating procedures
required by your organization
Figure 2 Spare part replacement flow
Item numbers
505000530
505000520
505000521
505000522
Order and receive
relevant parts to be
replaced.
Chapter 5:
Chapter 2:
Inspect the
Clean and disinfect
displaying unit
the displaying unit
7