Descargar Imprimir esta página

Eurosets HORIZON CVR Instrucciones De Uso página 35

Depósito de cardiotomía/depósito venoso para circulación extracorpórea en cardiocirugía

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 38
en
2) to it.
Connect the other end of the line to a vacuum
regulator equipped with a condensation trap.
Figure 2 – Connection line to the aspiration source
for VAVD
VENT/VACUUM
PORT
WARNINGS
• Do not apply the vacuum during priming and
recirculation procedure.
• It is recommended not to exceed -75.2 mmHg
(-10 KPa) when applying negative pressure to
the cardiotomy/venous reservoir.
• Periodically check functioning of the vacuum
regulating device and the degree of vacuum.
2) INITIATING BYPASS
OPENING THE ARTERIAL AND VENOUS LINES
If performing vacuum-assisted drainage, check
that all the connections are closed and apply the
vacuum to the cardiotomy/venous reservoir.
3) DURING BYPASS
CHECKING THE VENOUS RETURN
With vacuum-assisted drainage, the vacuum
applied to the cardiotomy/ venous reservoir must
be adjusted in order to obtain adequate venous
return.
LOW FLOW RECIRCULATION
In all cases that are expected a circulatory
arrest and the opening of a shunt between the
arterial and venous circuit, to maintain blood
on the move, it is necessary to stop the vacuum
application.
If is necessary to interrupt vacuum application,
remove one of the leakproof caps on the
cardiotomy/venous reservoir and remove clamp
from the collecting line to the suction source for
the VAVD technique.
Resume the arterial pump flow to the value
previously set. In case of vacuum-assisted
drainage, re-insert leakproof cap on the
cardiotomy/venous reservoir and re insert the
clamp on the collecting line to the suction source.
Gradually increase the negative pressure to
reach an adequate blood drainage.
CARDIOTOMY/VENOUS
MANAGEMENT
WARNING
• If you wish to change from vacuum-assisted
SUCTION
SOURCE
CLAMP FOR
VACUUM ON/OFF
CAUTION
• During the vacuum-assisted venous drainage
4) USING THE CARDIOTOMY/VENOUS RESERVOIR
FOR POST-OPERATIVE AUTOTRANSFUSION
(Vacuum)
Connect a line and a vacuum regulator to the
VENT/VACUUM port of the cardiotomy/venous
reservoir.
WARNING
• Periodically check functioning of the vacuum
5) HORIZON CVR REPLACEMENT
1) Suspend the application of vacuum to the
2) Disconnect all connections lines from the
3) Place a new HORIZON CVR on the holder.
6) MEDICAL DEVICES FOR USE WITH HORIZON CVR
When using the vacuum-assisted drainage
technique, it is recommended to use the
specific Eurosets connection lines between the
cardiotomy/venous reservoir and the vacuum
regulator.
drainage to conventional venous drainage,
suspend the vacuum application by detaching
the vacuum line so that the venous reservoir
is brought back to atmospheric pressure.
procedure pay attention to involuntary
opening of the Luer-locks or connectors
positioned on the turrets, as this causes a
reduction of the vacuum in the reservoir;
open the connectors only for the time
necessary for the operation required.
regulating device and the degree of vacuum.
cardiotomy/venous reservoir (when vacuum-
assisted drainage has been used).
HORIZON CVR.
Connect all the lines (venous line to the
cardiotomy/venous reservoir, pump segment
of the cardiotomy/venous reservoir to the
oxygenator, and possible vacuum line to the
cardiotomy/venous reservoir).
RESERVOIR