CISPR 11
Method: The AMN placed 0,8 m from the boundary of the unit under test and bonded to a ground
reference plane. This distance was between the closest points of the AMN and the EUT. All other units
of the EUT and associated equipment were at least 0,8 m from the AMN. All power was connected to
the system through Artificial Mains Network (AMN). Conducted voltage measurements on mains lines
Laboratory Parameters
Ambient Temperature
Relative Humidity
Fully configured sample scanned over
the following frequency range
Frequency (MHZ)
0,15 to 0,50
0,50 to 5
5 to 30
Supplementary information: The result in this table is the minimum margin to the limit.
Electromagnetic emissions
1.6 Test Conditions and Results - Conducted Emissions
TEST: Limits of mains terminal disturbance voltage
were made at the output of the AMN.
Required prior to the test
Frequency range on each side of line
150kHz to 30MHz
Limits - Group 1 and 2 - Class B - EUT Configuration mode 2
Quasi Peak
66 to 56
56
60
10 to 40 °C
10 to 90 %
Limit dB (uV)
Result *
12
7,1
16,8
90
During the test
22 °C
57%
Measurement Point
Mains
Average
56 to 46
46
50
Verdict
P
Result *
2
0,2
6,8