Table 3
Solvent
Acetone
Carbon tetrachloride
Chloroform
Chloroformmethanol (2:1)
Cyclohexane
Dichloromethane
Ethanol
Ethyl acetate
Heptane
Hexane
Methanol
Methyl Ethyl Ketone
Petroleumether 30-60
Toluene
Thrichloroethylene
Xylene
* = not recommended
The figures given in the table refer to aluminium and glass cups. OT (over-tempera-
ture) is the safety at out temperature if the temperature regulation would fail.
5.4 Operation
There are nine different programs.
Program #1 to #8 are locked. This means that when either of them has been started,
it is not possible to change temperature or time at any step.
During a run using program #9 it is possible to change temperature as well as time
setting. This program can be used to set the right temperature following the drop rate.
See section 5.4.5 Checking the Drop Rate on page 5:9.
5.4.1 Control Unit Settings
Press the "POWER" key on the Control Unit (switch should light up), and the display
will show the version number of the internal software before going to standby, e.g.
showing the previously run program.
5:8
Soxtec™ 2050 Automatic System
Temp
Overtemp
Al. Cups
Al. Cups
(°C)
(°C)
150
210
*
-
140
210
140-150
210
170
210
120-130
210
250
330
180
210
170-180
210
155
210
250-260
330
160
210
135
210
210-220
330
185
330
260
330
User Manual 1000 7414 / Rev. 4
Temp
Overtemp
Glass Cups
Glass Cups
(°C)
(°C)
180
330
185
330
170
330
170
330
*
-
160
210
285
330
210
330
180
210
180
210
285
330
220
330
160
210
285
330
240
330
285
330