Soxtec™ 2050 Automatic System
6.4.4 Control of Over-temperature 210
1.
Repeat the section 6.4.2 Control of Over-temperature 145 on page 6:2 and 5.4.9
Resetting the Over-temperature Protection on page 5:12 but select the over-tem-
perature 210.
2.
The power cuts when it reaches 190 degrees.
3.
If the temperature exceed 210 degrees, immediately contact FOSS Analytical
AB Service Department and do not use the instrument under any circumstances!
6.4.5 Control of Over-temperature 330
1.
Repeat the section 6.4.2 Control of Over-temperature 145 on page 6:2 and 5.4.9
Resetting the Over-temperature Protection on page 5:12 but select the over-tem-
perature 330.
2.
The power cuts when it reaches 290 degrees.
3.
If the temperature exceeds 330 degrees, immediately contact FOSS Analytical
AB Service Department and do not use the instrument under any circumstances!
6.5 Replacement
3
1
Fig. 6:1
Replacement of the solvent addition connectors
3
1
Fig. 6:2 Replacement of front glass
User Manual 1000 7414 / Rev. 4
6
4
2
5
4
2
5
2050034c
6
2050035c
6:3