Soxtec™ 2050 Automatic System
7 Trouble Shooting
7.1 Error Codes on Display
Error code
Explanation
Power failure during previous
ER 1
run
ER 2 + quick
Main temp higher than set over-
blinking
temperature
ER 3
Main temp too high
Calibration data for temperature
ER 4
sensor missing.
Difference between temperature
ER 5
sensors more than 30 degrees
Celsius.
Overtemperature protection
ER 6
measure below zero or the sen-
sor is disconnected.
Main temp measure below zero
ER 7
or the sensor is disconnected.
The pressure sensor is activated
ER 8
and the water pressure dissapear.
ER 80 or 81
Program failure
User Manual 1000 7414 / Rev. 4
Reset
Press START/STOP
Turn the power off, then on. The LED
is now blinking slowly. Turn the OT-
key. Alternatively check the sensor
connections.
Turn the power off and on. If the OT
cuts, turn the OT-key.
Contact the FOSS Analytical AB Serv-
ice Department
Turn the power off and on. If the OT
cuts, turn the OT-key.
Turn the power off and on. If the OT
cuts, turn the OT-key.
Turn the power off and on. If the OT
cuts, turn the OT-key.
Turn the power off and on. If the OT
cuts, turn the OT-key.
Contact the FOSS Analytical AB Serv-
ice Department