start a new event the dates must be configured in the same way described before to
cancel one event by date, configure the starting date later than the final date.
OBS.: the events by date have priority over the events by tension.
6.7. MEMORY, SAMPLING, REGISTRIES INTERVAL, MEMORY KIND.
This function configures the memory blocks, sampling, registries interval and
memory kind of the PowerNET P-600.
To enter in the memory configuration, press the key F0 to activate the second
function, when the display shows the letter f at the lower right corner press F4.
Registries blocks
The blocks define the measurements to be stored in one registry. Each block
has a determinated space in the memory, the more blocks enabled, the larger will be
the amount of memory used for registries. .
P"
The symbol "
will be stored inside the registry. When entering in memory configuration the screen
below will show:
TN
Configura Memoria = Configures Memory
Tensao = Tension
Corrente = Current
Potencia = Power
Energia = Energy
v Utilize the increase or decrease keys for enable or disable the block to be
registered. .
v Utilize the keys go to left or right to put the cursor over the desired block.
v To confirm and go to the next parameter press F4.
v To leave press F6.
To configure the maximum and minimum blocks and the harmonics block follow
the above mentioned procedure.
Sampling
This parameter defines the arithmetical average of the measurements each
200ms that the PowerNET P-600 will sample. For instance, if was configured a
sampling for 1000ms, the reading result of the measurements will be the average of 5
readings each 200ms.
PowerNET P-600
in the equipment display mean that the block is enabled and it
Installing and Operating Manual
122